이 사이트는 쿠키를 사용합니다. 사이트를 계속 탐색하시면 당사의 쿠키 사용에 동의하는 것입니다. 기본 설정은 사용자가 관리할 수 있습니다.
더 찾기

[MH680]Why there's no response when sending RS-232 commands with Extron Data viewer or Tera term to the projector?

03-06-2017
"It could be the connection between your system and the projector was not appropriately built. Please make sure the following items are functional -
1. The baud rate in projector OSD menu should be identical with the setting on Extron Dataviewer or Tera term. For example, to set baud rate as 115200 on the projector and baud rate on Extron Dataviewer or Tera term as 115200 too.
2. Make sure the RS-232 cable is ""Crossover"" type.
3. If using RS-232 with USB, make sure the USB driver is correctly installed. Also, the Com port should be correctly selected.
4. If connection was built, when press ""Enter"", you will see the projector echos a "">"" in Extron Dataviewer or Tera Term window."
TOP